DESCRIPTION
ADCA sample coolers are specially designed to cool samples of boiler water or steam for analysis. Sample coolers prevent steam flashing-off from hot pressurized liquid samples, which can be dangerous and may result in an incorrect water sample. This device may be used for boiler water analysis and other sampling or cooling applications compatible with the construction materials.
MAIN FEATURES
Corrosion-resistant body and internals.
Self draining sample (top inlet, bottom outlet).
STANDARD SURFACE FINISH
Internal wetted parts: ≤ 0,51 micron Ra – SF1.
External: ≤ 0,76 micron Ra – SF3.
Other surface conditions see IS PV20.00 E – Technical information.
Ultrasonic cleaning.
OPTIONS:
Different connection sizes and materials.
USE:
Steam boilers.
Hot water systems.
AVAILABLE MODELS:
SC32P – permanent sampling.
SC32PP – portable design.
SIZES AND CONNECTIONS:
Cooling water inlet/outlet: 1/2” on body (BSP or NPT) or 3/4" tri-clamp.
Sample tube inlet/outlet: 8 mm O/D. Alternative 1/2" tri-clamp compatible.
PACKAGING:
Assembling and packaging in a clean room certified according to ISO 14644-1. The product is end capped and sealed with recyclable thermo-shrinkable plastic film, to avoid contamination.
INSTALLATION:
Vertical installation.
OPERATION:
Cooling water must be at a maximum flow before opening or closing the sample inlet valve, in order to avoid scalding. The sample valve must also be closed before the cooling water valve.
PERFORMANCE:
30 to 60 kg/h of sample water at ≈ 30 oC with 1 m3 /h - 15 oC inlet cooling water (boilers up to 20 bar – 220 oC). For other pressures, temperatures and/or certified figures please consult.